Preface

THE PRODUCT

This manual describes changes/corrections implemented in the new version of the product "Test Programs For ND-100/110/120" (210523H00). The new version is mainly caused by changes in the SCSI-IV test program, but the TPE-MON-100-B, the DISK-MM-B, the CONFIGURATION-D, the FLOPPY-STREA-C, the POWER-FAIL-A, the NET-ONE-A, and the OCTOBUS-B programs also contain changes. See the PI-sheet Test programs for ND-100/110/120 (895076.1 EN) for information on the programs not described in this manual.

This product, and this manual, are intended for use only with computers having one, or more, of the devices 310MB or 155MB SCSI 5.25" disk drive (326392), the ND Gigatape System (326423), or the HP freestanding magtape drive (326363).

Document Description Code
Test Program Description for ND-100/ND-110 ND-830005.3 EN
Test Programs for ND-100/110/120 PI Sheet ND-895076.1 EN
BACKUP User Guide ND-860250.2 EN

Note

The numbering system for Norsk Data's documentation changed in September 1988. All numbers now start with an 8. The numbering structure is therefore ND-8xxxxxx.xx xx.
Example: ND-863018.3A EN. Existing manuals will receive a new number if and when they are updated or revised.

The information in this manual is subject to change without notice.
Norsk Data A.S assumes no responsibility for any errors that may appear in this manual, or for the use or reliability of its software on equipment that is not furnished or supported by Norsk Data A.S.

Copyright © 1988 by Norsk Data A.S
Version 1 December 1988

Send all documentation requests to:

Norsk Data A.S
Graphic Centre
P.O. Box 25 - Bogerud
N-0621 Oslo 6
NORWAY


Page 5

Table of Content

1 DISK-MM-B ............................................ 1

1.1 Purpose ........................................... 3
1.2 Target devices .................................... 3
1.3 Command description .............................. 3
1.4 Known, but not corrected errors ................... 7

2 NET-ONE-A ............................................ 9

2.1 Purpose .......................................... 11
2.2 Command descriptions ............................. 11
2.3 Known, but not corrected errors .................. 14

3 OCTOBUS-B ............................................ 15

3.1 Requirement ...................................... 17
3.2 Commands ......................................... 17
3.2.1 SET-PARAMETERS ................................. 18
3.2.2 SELECT-DEVICE .................................. 18
3.2.3 SELECT-OCTOBUS-STATION ......................... 19
3.2.4 LIST-HARDWARE-CONFIGURATION .................... 19
3.2.5 LIST-OCTOBUS-DEVICES ........................... 19
3.2.6 DECODE-STATUS-REGISTER .......................... 19
3.2.7 RUN ............................................ 20
3.2.8 OCTOBUS-FACILITIES ............................. 21
8.1 ACCESS-OCTOBUS-REGISTER ........................ 21
8.2 RESTART ....................................... 22
8.3 CONTINUE ...................................... 22
8.4 STOP .......................................... 22
8.5 INT7 .......................................... 22
8.6 RESET-COUNTER ................................. 22
8.7 POWER-UP ...................................... 22
8.8 POWER-DOWN .................................... 22
8.9 READ-OCTOBUS-TRANSMIT-STATUS ................... 22
8.10 RECEIVE-FROM-OCTOBUS ......................... 23
8.11 TRANSMIT-ON-OCTOBUS .......................... 23
8.12 SELECT-DEVICE ................................ 23

4 SCSI-TV-C ............................................ 25

Scanned by Jonny Oddene for Sintran Data © 2012


Page 6

Table of Contents

4.1 Products supported by version C 27
4.2 Command description 27
4.3 Host adapter tests 30
4.4 Device tests 32
4.5 Known but not Corrected Errors 35

Scanned by Jonny Oddene for Sintran Data © 2012


Page 7

Test Program Description for ND-100/ND-110 - Addendum

CHAPTER 1

DISK-MY-B


Scanned by Jonny Oddene for Sintran Data © 2012


Page 8

Test Program Description for ND-100/ND-110 - Addendum


[No visible diagrams or tables on this page]


Scanned by Jonny Oddene for Sintran Data © 2012


Page 9

DISK-MM-B

1.1 Purpose

The DISK Media Maintenance program is used to format, initialize and verify SCSI disks connected to the ND-100 SCSI host adapter.

1.2 Target devices

The program is designed to handle the

Micropolis 1375 SCSI disk drive.
CDC EMD 97201 SCSI disk drive.

1.3 Command description

›SELECT-DEVICE

This command will select a device for use by other commands. Commands that use a device for operations will ask for it if it is not defined by this command.

Legal answers for adapter are:

  • adapter name (adapter-n, n=1:4).
  • adapter number (1:4).
  • Logical device number for adapters data field {2202B:2205B}.
  • Hardware device number for adapter {144300B, 144400B, 144500B, 144600B}

Legal answers for ID number are 0:7. ID number 7 is normally the SCSI host adapter, and ID 0 is normally equivalent to "DISC-SCSI-1" which is the boot disk.

The default adapter is "adapter-1" or the last selected adapter, and the default ID is "0" or the last selected ID number.

›CONVERT-DISK-ADDRESS

This command will convert a disk to different address modes and numeric systems.

The address type could be "Media", "Logical-SINTRAN" or "page".

Scanned by Jonny Oddene for Sintran Data © 2012


Page 10

Test Program Description for ND-100/ND-110 - Addendum

DISK-MM-B

Media Address

A media address is the address used on basis of the disk drive's sector size.

Logical-SINTRAN

A logical-SINTRAN address is the address used on basis of SINTRAN's block size, which always is 1024 bytes/sector.

Page

A page address is the address used on basis of SINTRAN filesystem's block size, which always is 2048 bytes/sector.

FORMAT-INITIALIZE

Formats and initializes the disk.

Special for Micropolis 1375:

Before format starts, MODE-SENSE default page 3 and 4 is set to current, and cylinder skew and track skew is optimized.

Operation Times

Disk Type Size Time Used (approx.)
Micropolis 1375 60Mb 2:00 Min.
Micropolis 1375 125Mb 4:00 Min.
CDC EMD 97201 310Mb 10:00 Min.
CDC EMD 97201 630Mb 20:00 Min.

INITIALIZE

Prepares the disk for use under SINTRAN.

LIST-DISK-INFORMATION

This command will print out various static and dynamic information concerning the selected disk.

LIST-PARAMETERS

This command will print out the setting of all parameters available in the command "SET-PARAMETERS".

LIST-REASSIGN-TABLE

This command will print out all sectors reassigned with the command "REASSIGN". The sectors are listed in historical order.

LIST-REFRESH-TABLE

This command will print out all sectors refreshed with the command "REFRESH-DATA". The sectors are listed in historical order.


Scanned by Jonny Oddene for Sintran Data © 2012


Page 11

Test Program Description for ND-100/ND-110 - Addendum

REASSIGN <Media address>

This command will reassign one single sector and its data on the disk to a different physical area, but to the same logical area.

The <Media address> parameter is the disk address of the sector containing a flaw. The address could be in any partition.

Reassigning a sector means to relocate it physically or the entire physical track. This is done by the disk drive itself, and the program has no control over physical actions on media. After the reassign operation, the physical area earlier used for the sector will no longer be used.

On successful reassign, the reassign operation will be logged in the table partition and can later on be listed with the command "LIST-REASSIGN-TABLE".

If the program is not able to read the sector addressed, a confirmation question has to be answered.

REFRESH-DATA <Media address>

This command will try to read one sector and write it back trying to cure temporary parity errors.

The <Media address> parameter is the disk address of the sector to be read and written. The address could be in any partition.

If the physical space on media holding the sector contains flaws, it will probably not be possible to refresh a sector. In such cases, a medium error will be reported, and it is advised to reassign the sector.

On successful refresh, the refresh operation will be logged in the table partition and can later on be listed with the command "LIST-REFRESH-TABLE".

If the program is not able to read the sector addressed, a confirmation question has to be answered.

SET-PARAMETERS <Parameter> <Value>

This command can be used for setting of special parameters for device operation.

The parameters are:

  • BUS-RESET

    This parameter decides whether a SCSI bus reset can be done by the program or not.

    In multi-host systems, this parameter should be set to "No" if the other hosts are running.


Scanned by Jonny Oddene for Sintran Data © 2012


Page 12

Test Program Description for ND-100/ND-110 - Addendum

DISK-MM-8

RESET-TIMEOUT

This parameter sets the timeout after bus reset. The timeout must be larger than the time needed for disk power up selftest to complete. This parameter is only significant if the "BUS-RESET" is "Yes".

ADDRESS-INPUT-RADIX

This parameter sets the radix for disk address inputs. Possible values are "Octal", "Decimal", "Binary" and "Hexadecimal". The program has initially default "Decimal".

FIRST-PHYSICAL-BUFFER-PAGE

This parameter sets the first page in physical memory that should be used by the program for all DMA access to the selected adapter.

LAST-PHYSICAL-BUFFER-PAGE

This parameter sets the last page in physical memory that should be used by the program for all DMA access to the selected adapter.

VERIFY

Data verification.

  • The addresses are given in media blocks.
  • Default range: <0,lastBlock>.

Operation Times

Disk type Size Time used (approx.)
Micropolis 1375 60Mb 1:00 Min.
Micropolis 1375 125Mb 2:00 Min.
CDC EMD 97201 310Mb 3:30 Min.
CDC EMD 97201 630Mb 7:00 Min.

Page 13

Test Program Description for ND-100/ND-110 - Addendum

1.4 Known, but not corrected errors

  • The program does not clean properly up after errors messages from MPM4.

    If the error message:

    MPM4 Memory out of range
    

    appear, reload the program and use the commands:

    >SET-PARAMETERS,FIRST-PHYSICAL-BUFFER-PAGE,<MPM5 first page>
    >SET-PARAMETERS,LAST-PHYSICAL-BUFFER-PAGE,<MPM5 last page>
    

    for setting of correct buffer limits before selecting the device.

    If one of the error messages:

    MPM4 Parity error
    MPM4 Power fail
    

    appear, please reload the program before you continue.


Page 14

8       Test Program Description for ND-100/ND-110 - Addendum


Scanned by Jonny Oddene for Sintran Data © 2012

Page 15

Test Program Description for ND-100/ND-110 - Addendum

Chapter 2

NET-ONE-

Scanned by Jonny Oddene for Sintran Data © 2012


Page 16

Test Program Description for ND-100/ND-110 - Addendum

[Page is blank]


Scanned by Jonny Oddene for Sintran Data © 2012


Page 17

Test Program Description for ND-100/ND-110 - Addendum

2 NET-ONE-A

2.1 Purpose

This program tests the NET/ONE interface controller, also referred to as the NOTS controller.

2.2 Command Descriptions

>SELECT-DEVICE

Selects NOTS controller to be tested (1:8).

>LIST-ALL-DEVICES

Prints a list of the defined NOTS controllers.

>LIST-DEVICES-PRESENT

Prints a list of all present NOTS controllers.

>RUN

This command makes it possible to execute one specific test, or a subset of all tests in a specified sequence.

Tests available:

Test Number Test Name
1 REGISTER-IOX TEST
2 TIMER-IDENT TEST
3 MEMORY-PATTERN TEST
4 MEMORY-ADDRESS TEST
5 MBNIU-DIAGNOSTIC-1
6 MBNIU-DIAGNOSTIC-2

Test 1: Register-iox test

Tests the NOTS registers, i.e., after writing to a control register, the corresponding status register must contain the expected value.

Test 2: Timer-ident test

Tests the NOTS timer. The returned ident code must have the expected value.

Test 3: Memory-pattern test

Tests the MBNIU memory by writing known values, read back and compare.


Page 18

Test Program Description for ND-100/ND-110 - Addendum

Test 4: Memory-address test

This test uses a 'address in address' pattern when writing to the MBNIU memory, detects memory addressing errors.

Test 5: MBNIU-diagnostics 1

This test runs the self-test diagnostic number 1 on the MBNIU board.

Test 6: MBNIU-diagnostics 2

This test runs the self-test diagnostic number 2 on the MBNIU board. (It takes about 4 minutes to run this test)

MBNIU: MultiBus Network Interface Unit.

Note: The error messages reported from Test 5 and Test 6 are a copy of the LED pattern on the MBNIU card.

SET-PARAMETERS

Defines how the program should run the tests.

Syntax:

SET-PARAMETERS <loop mode {<loops>}
               <abort mode {<errors>}
               <suppress mode>
               <debug mode>
Parameter Description Default
<loop mode> YES or NO. YES means the test or tests is repeated. NO means that the test is run only once. NO
<loops> Number of times the test is to be repeated. Infinite
<abort mode> YES or NO. YES means that the test(s) is aborted when maximum number of errors is reached. NO means that the test(s) will never abort. NO
<errors> Maximum errors allowed before abortion. 10

Scanned by Jonny Oddene for Sintran Data © 2012


Page 19

Test Program Description for ND-100/ND-110 - Addendum

NET-ONE-Acrev

<suppress mode> : YES or NO
YES means that the error messages are suppressed.
NO means that the error messages are printed.
Default : NO

<debug mode> : YES or NO
YES turns the debug mode on.
NO turns the debug mode off.
Default : NO

NOTS-DEBUG

This command is only available when the debug mode is on (see SET-PARAMETERS). The command has several subcommands which are listed below.

List-debug-parameters

Shows the current debug parameters.

Timer-units

Changes the timer value used for the Timer/Ident test.

Address-range

Specifies from/to addresses used when testing the NOTS memory.

Edit-test-patterns

Makes it possible to change the test patterns used by the memory pattern test.

Operations
CR display next pattern
<pattern number>/ jump to specified pattern number
<value> (cr) change pattern
space delete entry
. (point) exit

Modifier (Yes/No)

When turned on the memory address test will add a displacement factor when computing the test pattern.

Look-at-NOTS-memory <bank number>

The user may examine and change contents of memory locations within one bank.


Page 20

Test Program Description for ND-100/ND-110 - Addendum

NET-ONE-A‹rev›

Operations

Operation Description
CR display contents of next memory location
<address>/ jump to specified address
<value> (cr) change contents of current memory location
space delete entry
lower\ dump memory block
, (point) exit

2.3 Known, but not corrected errors

Test 2 will fail if Timer-Units (see NOTS-DEBUG) are greater than 7.


Page 21

Test Program Description for ND-100/ND-110 - Addendum

Chapter 3

OCTOBUS-B

[Scanned by Jonny Oddene for Sintran Data © 2012]


Page 22

Test Program Description for ND-100/ND-110 - Addendum

[Page content is empty]


Scanned by Jonny Oddene for Sintran Data © 2012


Page 23

OCTOBUS-B

The Octobus Test Program runs stand alone in the ND-100, controlled by the TPE-monitor. The program must be downloaded from floppy. Its basic functions are:

  • Test the Octobus controller in the ND-100 Line Driver.
  • Test the Octobus communication between the Octobus controller in the ND-100 Line Driver and the present Domino controllers.
  • The Octobus communication between the ND-100, the MFBus controller and the ACCP can be tested manually by using the Octobus Test Program, the MFBus Test and Maintenance program and the ACCP Console Monitor.
  • Find the Octobus configuration.

3.1 Requirement

Following requirements must be fulfilled to run the program satisfactory:

  • The configuration of the Octobus must follow the specification: A Domino controller may have Octobus station number from 10b to 67b.
  • Test 4, test 5 and test 6 require Domino prom version not older than version C (Domi-Opcom, 73100C).

3.2 Commands

The commands available for the user are:

  • SET-PARAMETERS \ \ \ \ \ \
  • SELECT-DEVICE \
  • SELECT-OCTOBUS-STATION \
  • LIST-HARDWARE-CONFIGURATION
  • LIST-OCTOBUS-DEVICES
  • DECODE-STATUS-REGISTER \ \
  • RUN \
  • OCTOBUS-FACILITIES

Scanned by Jonny Oddene for Sintran Data © 2012


Page 24

Test Program Description for ND-100/ND-110 - Addendum

3.2.1 SET-PARAMETERS

The user may set parameters which decide the behaviour of the RUN command. The parameters are listed below, and their default values are shown in parenthesis.

  • Loop mode (No)
  • Abort mode (Yes)
  • After how many errors (10)
  • Suppress error messages (No)
  • Define error reporting level (No)
  • Test all present Octobus devices (No)
  • Maximum message length (bytes) (255)

The parameter 'Maximum message length' specifies the size of the largest multibyte message echoed between the ND-100 and a Domino controller in test 6 (Echo multibyte message). Cannot be greater than 255.

The user may specify the appearance of the error messages. This is done by answering 'Yes' to the question 'Define error reporting level'. The following questions must then be answered (default values in parenthesis):

  • Controller number (Yes)
    The number of the failing Octobus Controller.
  • Hardware device number (Yes)
    The hardware device number for the failing Octobus controller.
  • Type of error (Yes)
    Specify what is wrong.
  • Error information (Yes)
    This information depends on type of error. It may be register contents, or found and expected values.
  • Decoding of status (Yes)
    Decoding of register contents shown under 'Error information'.

3.2.2 SELECT-DEVICE

The user has the possibility to choose which Octobus device to test, if there are more than one Octobus controller present. Default is that the device with the lowest Octobus controller number is tested.


Page 25

Test Program Description for ND-100/ND-110 - Addendum

OCTOBUS-B

If one wants to test all present Octobus devices, this may be specified under SET-PARAMETERS. Default is that only one Octobus controller is tested.

3.2.3 SELECT-OCTOBUS-STATION

In test 4, 5 and 6 the Octobus communication between ND-100 and the Domino controllers is tested. Default is that the communication between ND-100 and all possible Octobus stations is tested. If you want to test the communication between ND-100 and one specified Octobus station, you have to specify this station by using this command.

3.2.4 LIST-HARDWARE-CONFIGURATION

Will find present Octobus controllers, and all the Octobus stations present for each controller. These stations may be the MFBus controller, the ACCP and/or the Domino modules. Before returning to TPE, a configuration table is shown. This table consists of, from left to right:

  • Octobus controller number.
  • Octobus hardware device number.
  • Receive ident code (level 13).
  • Transmit ident code (level 13).
  • The Octobus controller's station number.
  • Stations seen by the Octobus controller.

3.2.5 LIST-OCTOBUS-DEVICES

This command presents a table showing all devices defined for the ND-100 Octobus interface, by Norsk Data. This table consists of, from left to right:

  • Octobus controller number.
  • Octobus hardware device number.
  • Receive ident code (level 13).
  • Transmit ident code (level 13).

3.2.6 DECODE-STATUS-REGISTER

It presents a decoding of a given Octobus status register. The status register may be transmit or receive.


Page 26

Test Program Description for ND-100/ND-110 - Addendum

3.2.7 RUN

Using this command, you can run all tests (default), only one test or a sequence of tests. The available tests are:

  1. Check transmit - receive loop.
  2. Loop all possible patterns.
  3. Check receive fifo length.
  4. Check Octobus configuration.
  5. Echo single word messages.
  6. Echo multi word messages.

Test 1: The controller will send one byte to itself. The transmit and receive parts are tested.

Test 2: The controller will send all possible bit patterns to itself, and compare the transmitted and received patterns. It is tested whether the controller is able to transmit and receive all possible bit patterns in different order.

Test 3: The controller will send several bytes to itself, and detect when the receive fifo is full. The size of the receive fifo is checked.

Test 4: The controller communicates with the present Octobus stations which are Domino controllers, via the Octobus. Each Domino controller answers an "Identify-yourself" message. The responses are compared with the hardware configuration list found without activating the Octobus stations. Each Domino controller is then asked to present a list showing the Octobus stations it "sees". All the incoming lists are compared to see whether "all can see all".

Test 5: The controller communicates with the present Octobus stations which are Domino controllers, via the Octobus. Small multibyte messages (8 bytes) containing a varying word pattern are echoed between the controller and the Domino controllers.

Test 6: The controller communicates with the present Octobus stations which are Domino controllers, via the Octobus. Multibyte messages with varying lengths are echoed between the controller and the Domino controllers.

The program will always test the following items, without user intervention:

  • Status registers.
  • Interrupt and ident codes.

Page 27

Test Program Description for ND-100/ND-110 - Addendum

OCTOBUS-8

  • The combination RFT (Ready for Transfer), IE (Interrupt Enabled) and ID (Interrupt Detected).

3.2.8 OCTOBUS-FACILITIES

New command level. It gives the user the possibility to use the Octobus manually.

Corresponding commands exist in the MFBus Test and Maintenance program and in the ACCP Console Monitor. The Octobus can be tested manually by connecting a terminal via ASYL to the MFBus controller and the ACCP, and sending messages between the ND-100 Line Driver, the MFBus controller and the ACCP.

In addition there are some commands sending emergency messages which are decoded by the hardware at the destination Octobus station.

Following commands are available:

  • ACCESS-OCTOBUS-REGISTER <function> <register contents>
  • RESTART <dest>
  • CONTINUE <dest>
  • STOP <dest>
  • INT7 <dest>
  • RESET-COUNTER <dest>
  • POWER-UP <dest>
  • POWER-DOWN <dest>
  • READ-OCTOBUS-TRANSMIT-STATUS
  • RECEIVE-FROM-OCTOBUS <loop>
  • TRANSMIT-ON-OCTOBUS <dest> <control> <broadcast> <no of bytes> <byte no 1> .. <byte no n>
  • SELECT-DEVICE <Octobus controller number>

3.2.8.1 ACCESS-OCTOBUS-REGISTER

The user has direct access to all the Octobus registers. The available functions are:

  • READ-RECEIVE-DATA
  • READ-RECEIVE-STATUS

Page 28

Test Program Description for ND-100/ND-110 - Addendum

OCTOBUS-8\<rev>

  • WRITE-RECEIVE-CONTROL
  • WRITE-TRANSMIT-DATA
  • READ-TRANSMIT-STATUS
  • WRITE-TRANSMIT-CONTROL

Dependent of the function (read or write), the register content is presented to the user, or the user must specify the content.

3.2.8.2 RESTART

Hardware decoded message to specified Octobus station. Activates the RESET signal and restarts the controller after a total reset.

3.2.8.3 CONTINUE

Hardware decoded message to specified Octobus station. Deactivates the HALT signal.

3.2.8.4 STOP

Hardware decoded message to specified Octobus station. Activates the HALT signal. Halt must remain active until the CONTINUE message is received.

3.2.8.5 INT7

Hardware decoded message to specified Octobus station. Generates a level 7 interrupt. Force the processor out of a hang situation.

3.2.8.6 RESET-COUNTER

Hardware decoded message to specified Octobus station. Resets the time reference counter.

3.2.8.7 POWER-UP

Hardware decoded message to specified Octobus station. Power up.

3.2.8.8 POWER-DOWN

Hardware decoded message to specified Octobus station. Power down.

3.2.8.9 READ-OCTOBUS-TRANSMIT-STATUS

The content of the Octobus transmit status register is presented to the user, and decoded.


Scanned by Jonny Oddene for Sintran Data © 2012


Page 29

Test Program Description for ND-100/ND-110 - Addendum

OCTOBUS-Brev

3.2.8.10 RECEIVE-FROM-OCTOBUS

If the Octobus receive fifo is not empty, one byte from the fifo is presented to the user. The presentation is in table form, one column showing the transmitter of the message, and one column showing the byte transmitted. It is possible to loop this command.

3.2.8.11 TRANSMIT-ON-OCTOBUS

It is possible to transmit maximum five bytes on the Octobus at a time. The user must specify destination station number, control- and broadcast-bit, how many bytes to transmit and their content.

3.2.8.12 SELECT-DEVICE

The behaviour of this command is the same as for the SELECT-DEVICE command of the above command level (TPE command level).


Page 30

Test Program Description for ND-100/ND-110 - Addendum


[Scanned by Jonny Oddene for Sintran Data © 2012]


Page 31

Test Program Description for ND-100/ND-110 - Addendum

Page 25

CHAPTER 4

SCSI-TV-C

[Scanned by Jonny Oddene for Sintran Data © 2012]


Page 32

[Page 32 — not captured by OCR (blank, image-only, or unreadable in the scan).]

Page 33

Test Program Description for ND-100/ND-110 - Addendum

4 SCSI-TV-C

Test program for the Floppy/SCSI print (SCSI part only) and SCSI devices connected to the bus.

This program is designed to run STAND-ALONE as well as under SINTRAN (ver. L) as user SYSTEM.

4.1 Products supported by version C

Vendor Product Device type
NDMICROP 1375 Direct
TANDBERG TDC 3600 Sequential (streamer)
OSI LD 1200 SCSI Write Once (optical disk)
NDCDC EMD 97201 (736) Direct
NDCDC EMD 97201 (368) Direct
NDCDC 94171-9 Direct
NDSTK 2925 Sequential (magtape)
ARCHIVE VIPER 150 21247 Sequential (streamer)
Exabyte 8200 Sequential (streamer)
HP 88780 Sequential (magtape)

4.2 Command description

  • CLEAR-DEVICE

    This command will perform a bus reset on the selected SCSI bus. Can be used in "hang" situations.

  • CREATE-OPTICAL-TEST-DISK

    This command makes a test disk for use in the optical disk tests. The last block on the disk is used as 'optical disk control record'. The first record on the disk is written with the correct test pattern.

    Note: The disk cartridge cannot be used for other purposes after being created as a test disk.

  • DECODE

    This command can decode some SCSI items. These are:

    • INTERFACE-REGISTER

      Decodes the different bits and codes in the ND-100 interface.

    • MESSAGE-CODE

      Decodes the SCSI message code.

Scanned by Jonny Oddene for Sintran Data © 2012


Page 34

Test Program Description for ND-100/ND-110 - Addendum

SCSI-TV-C

EXTENDED-MESSAGE-CODE <Value>

Decodes the extended SCSI message code.

COMMAND-CODE <Value>

Decodes the SCSI command operation code.

SENSE-KEY <Value>

Decodes the SCSI sense key.

ADDITIONAL-SENSE-CODE <Value>

Decodes the SCSI extended sense key.

STATUS-CODE <Value>

Decodes the SCSI status byte.

MEMORY-ADDRESS <Value>

Decodes a physical memory address in bank, page in bank, and displacement in page.

Commands

>DUMP-SENSE-DATA

Gives a hex dump of the contents of the sense block from the last request sense (i.e., the last error message reported from the SCSI device). The operator could then use the reference manual to decode the different field in the sense block.

>LIST-DEFINED-DEVICES

This command lists all SCSI devices defined under SINTRAN.

>LIST-PRESENT-DEVICES

This command will list all SCSI host adapters in ND-100 and the connected devices along with some important device information.

Prints the statistic data which are collected while running the magtape/streamer tests. (Only available when a sequential device is selected).

>RUN-adapter-TESTS <Test list>

This command will execute all or a specified sequence of tests for the ND-100 SCSI host adapter.

>RUN-DEVICE-TESTS <Test list>

This command will execute all or a specified sequence of tests for the SCSI device (e.g., disk or streamer).


Page 35

Test Program Description for ND-100/ND-110 - Addendum

SELECT-DEVICE <adapter name> <Id number> / <Device name>

You use this command to select the device to test. The command is used in two ways:

  1. In stand-alone mode where you give the host adapter number and SCSI id.
  2. In SINTRAN mode where you give the logical device name for the device.

Operating stand-alone:

Use SCSI host adapter and the SCSI ID number to select device.

Legal answers for adapter are:

  • Adapter name (adapter-n, n=1:4).
  • Adapter number (1:4).
  • Logical device number for adapters data field (22028:22058).
  • Hardware device number for adapter (144300B, 144400B, 144500B, 144600B).

Legal answers for SCSI ID number are 0:7. ID number 7 is normally the SCSI host adapter, and ID number 0 is normally equivalent to "DISC-SCSI-1" which is the boot disk.

Operating under SINTRAN:

Use logical device name defined under SINTRAN. <HELP> will give you all legal device names. The device name can be abbreviated according to SINTRAN abbreviation rules. "DISC-SCSI-1" is normally the boot disk.

SET-BUFFER-LIMITS

The operator may set first and last memory page for the DMA buffer (only used when the host adapter is placed in MULTIPORT-4).

SET-OPTICAL-DISK-PARAMETERS

This command allows you to set some parameters for the optical disk tests.

SET-PARAMETERS <Loop mode> <Loop count> <Abort mode> (<Abort count>) <Supress mode>

This command will set parameters for execution of the commands "RUN-adapter-TESTS" and "RUN-DEVICE-TESTS".

SET-TAPE-PARAMETERS

Change the default parameters for the write/read magtape test (test number 20) and the write/read streamer test (test number 30). (Only


[Scanned by Jonny Oddene for Sintran Data © 2012]


Page 36

Test Program Description for ND-100/ND-110 - Addendum

SCSI-TV-C

available when a sequential device is selected).

TAPE-SERVICES

This command performs some useful magtape/streamer services (LOAD, UNLOAD and REWIND). (Only available when a sequential device is selected).

PROGRAM-STATUS (TPE command)

This will include the values of some flags in the program.

4.3 Host adapter tests

Adapter tests available:

  1. Registers write/read back
  2. Control/Status
  3. DMA in test mode
  4. Interrupt at level 11
  5. NCR chip self diagnostic test
  6. NCR chip SCSI commands/status

Test 1: Registers write/read back

All interface registers which may be written and read back without triggering any special logic functions are tested with several bit patterns. The registers are:

  • Memory address
  • Data
  • SCSI data
  • SCSI control
  • Destination ID
  • Transfer counters

Test 2: Control/Status

The interface status registers is checked when the following bits are activated and reset in the interface control register:

  • Bit 0: Enable interrupt
  • Bit 2: Activate
  • Bit 4: Clear device
  • Bit 10: SCSI bus reset

Scanned by Jonny Oddene for Sintran Data © 2012


Page 37

Test Program Description for ND-100/ND-110 - Addendum

SCSI-TV-C

Test 3: DMA in test mode

The interface's test mode is used for DMA transfers. The test sequence is like this:

  1. Basic check on some of the control register bits (Test mode(Bit 3), DMA enable(Bit 5), Write enable(Bit 6)).
  2. Dynamic memory address register test.
  3. Several DMA transfers to and from memory.

Test 4: Interrupts at level 11

The test verifies that interrupt on Read For Transfer behaves ok with the correct IDENT code on level 11. It also verifies that the NCR chip interrupt is detected, and that this interrupt in turn will generate the interface interrupt.

Test 5: NCR chip self diagnostic test

The self diagnostics results on a device reset are verified at the NCR chip level, and the data turnaround tests are executed and verified with different patterns.

Test 6: NCR chip SCSI commands/status

Some basic functions for the NCR chip are verified, as well as the internal registers:

  • Pause bit in the auxiliary status.
  • Illegal select or reselect function using own ID.
  • All commands which are illegal in the disconnected state.

Page 38

4.4 Device Tests

Device tests available:

  1. Drive contact
  2. Basic drive self test
  3. Extended drive self test
  4. SCSI bus data transport
  5. SCSI bus stress

    ┌──────────────────────────┐
    │ Device controller tests  │
    └──────────────────────────┘
    
  6. Seek min/max

  7. Random seek
  8. Write/read scratch
  9. Random read
  10. Random read data partition - write scratch

    ┌────────────────────────┐
    │ Disk tests             │
    │ (none destructive)     │
    └────────────────────────┘
    
  11. Write/read magtape

  12. BOT test
  13. Wear test
  14. Filemark test

    ┌──────────────┐
    │ Magtape tests│
    └──────────────┘
    
  15. Write/read streamer

  16. Over/under run
  17. Filemark/space test

    ┌────────────────┐
    │ Streamer tests │
    └────────────────┘
    
  18. Write to optical

  19. Read/test optical

    ┌────────────────────┐
    │ Optical disk tests │
    └────────────────────┘
    

Note: The NDSTK 2925 magtape drive must be in 'select mode' (selected with the density key on the operator panel) when running the tests 20:23.

Test 1: Drive contact

A SCSI inquiry command is executed on the selected drive to verify contact with the drive and to test its identification.

NB This test will disable other tests in the test link if the drive found does not support some commands.

NB This test is always executed initially in every device test.

Test 2: Basic drive self test

This test uses the SCSI command "Send Diagnostic" with only the self test bit set. For most drives, this will activate the power-up self test.

Test 3: Extended drive self test

This test activates a more extended self-test in the drive (if available).


Page 39

Test Program Description for ND-100/ND-110 - Addendum

Test 4: SCSI bus data transport

This test will verify the SCSI bus data transport to the selected drive. It uses the SCSI commands "Write Data Buffer" and "Read Data Buffer". The drive buffer length is found and the test sequence is repeated with different test patterns. The test sequence is:

  • Write a test pattern to the buffer.
  • Clear the ND-100 memory buffer.
  • Read data back from drive.
  • Test the data and report errors.

The test patterns used are:

  • All bits zero.
  • All bits set.
  • Walking bit set.
  • Walking bit cleared.
  • One byte zero and next byte with all bit set.
  • Every second bit set.
  • Every second bit set in one byte and the opposite bits set in the next byte.

Test 5: SCSI bus stress

This test is similar to test 4 except that the data read back is not tested, and the same buffer is read 40 times.

Test 10: Seek min/max

Performs continuous seek between address zero and the (logical) last address on the selected disk.

Test 11: Random seek

Performs random seek on the whole disk.

Test 12: Write/read scratch

Writes different patterns with different transfer lengths to the scratch partition on the disk.

Test 13: Random read

Performs random read on the whole disk.

Test 14: Random read data partition - write scratch

Reads data from the data partition and writes it to the scratch partition. The scratch block and the original block in the data partition.

[Scanned by Jonny Odden for Sintran Data © 2012]


Page 40

Test Program Description for ND-100/ND-110 - Addendum

Test 20: Write/read magtape

This test simulates normal use of the magtape drive (see also SET-TAPE-PARAMETERS).

Test 21: BOT test

This test checks correct positioning at BOT for various rewind conditions.

Test 22: Wear test

The purpose of this test is to try to detect a deterioration due to repeated writing and reading.

Test 23: Filemark test

The purpose of this test is to check the detection of a filemark under various circumstances.

Test 30: Write/read streamer

This test simulates normal use of the drive (see also SET-TAPE-PARAMETERS).

Test 31: Over/under run

Writes and reads data with delay between the next data transfer to provoke over/under run.

Test 32: Filemark/space test

Writes and reads data with different number of records for each operation, after a write operation a filemark is written. Space over filemark, space over records and space to end-of-recorded-data are also tested.

Test 40: Write to optical

Writes a specified number of blocks to the optical disk. This test may operate in two different modes:

  1. The standard test area, which is the last megabyte on the disk, will not be used by other ND software products and is therefore used as default by SCSI-TV for write and read on the disk media.

  2. If you use a test prepared with the CREATE-OPTICAL-TEST-DISK command, the whole disk will be reserved for the media tests.

Note: The write enable parameter in the SET-OPTICAL-DISK-PARAMETERS must be set to YES.

Test 41: Read/test optical disk

Read back and test the last blocks written by the write to optical disk test.


Page 41

Test Program Description for ND-100/ND-110 - Addendum

4.5 Known but not Corrected Errors

  • Early versions of the MICROPOLIS disk 1370 for 60MB will not be recognized by the test program.
  • The program does not clean properly up after errors messages from MPM4.

    If the error message:

    MPM4 Memory out of range
    
    appear, reload the program and use the commands:

    >SET-BUFFER-LIMITS <MPM5 first page>,<MPM5 last page>
    
    for setting of correct buffer limits before selecting the device.

    If one of the error messages:

    MPM4 Parity error
    MPM4 Power fail
    
    appear, please reload the program before you continue.

  • The program fails if there is too little memory available for swapping in SINTRAN.

    If the following message:

    *E* (050353B) (PB) No memory available
    
    appears when loading the program, use the ND-500/5000 MONITOR to take some memory from the 500(0) processes.

    Use the following commands before reloading the program:

    @ND-500
    N500: TAKE 100
    N500: EXIT
    

Page 42